Table 2 Measurement differences between the ViMove and Vicon systems

From: Clinically acceptable agreement between the ViMove wireless motion sensor system and the Vicon motion capture system when measuring lumbar region inclination motion in the sagittal and coronal planes

Inclination direction

Root mean squared error (SD)

Mean difference

Lower limit of agreement

Upper limit of agreement

Flexion

1.82 ± 1.00

0.42

-3.86

4.69

Extension

0.71 ± 0.34

-0.12

-2.15

1.91

Lateral Right

0.77 ± 0.24

-0.16

-2.37

2.05

Lateral Left

0.98 ± 0.69

-0.08

-3.11

2.96

  1. SD standard deviation (SD). All values are presented in degrees of movement