Skip to main content

Table 4 Humeral implant related complication

From: Anatomic stemless shoulder arthroplasty and related outcomes: a systematic review

   

Implant

n

Percentage of complication

Kind of stemless humeral implant related complication and treatment

2010

Huguet et al.

JSES

TESS

63

7.9 %

- Five patients with a small crack of the humeral lateral cortex intraoperatively, noticed on the first postoperative radiograph, further conservative treatment

2013

Razmjou et al.

JSES

TESS

17

0 %

-

2013

Berth et Pap

JOT

TESS

41

0 %

-

2015

Maier et al.

BMC

TESS

12

0 %

-

2011

Kadum et al.

AOTS

TESS

22

0 %

-

2014

Bell et Coghlan

Int J Shoulder Surg

Mathys Affinis

50

0 %

-

2011

Schoch et al.

Obere Extremitaet

Arthrex Eclipse

115

0 %

-

2012

Brunner et al.

Obere Extremitaet

Arthrex Eclipse

233

2.3 %

- One patient with radiological and asymptomatic loosening after 24 months

2015

Habermeyer et al.

JSES

Arthrex Eclipse

78

0 %

-

2016

Ho et al.

JSES

Simpliciti

149

0 %

-

2016

Churchill et al.

JBJS

Simpliciti

149

0 %

-