Table 2 Difference in degrees between the true limitation (measured with EMTS) and the limitation detected by the simulation software

From: A quantitative non-invasive assessment of femoroacetabular impingement with CT-based dynamic simulation - cadaveric validation study

Hip joint 1 2 3 4 5  
ci R R L R L Median
Max flexion −2 −1 −2 −2 −4 −1.7
Max abduction −1 0 1 −1 −1 −0.5
Max internal rotation 0 3 −7 −2 −1 −0.8
Max internal rotation at 30° 0 −1 2 −1 1 0.1
Max internal rotation at 60° −1 −1 5 −2 −5 −1.3
Max internal rotation at 90° 02 1 0 −3 5 0.0