Skip to main content

Table 1 Intra-Rater ICC for all human evaluators and the computer program. There was increased reproducibility, in general, with increased experience. EX3 demonstrated the highest ICC (0.99) while the computer program was perfectly reproducible (ICC = 1.00).

From: Validation and reproducibility of computerised cell-viability analysis of tissue slices

Evaluator

ICC

US1 – unskilled

0.75

US2 – unskilled

0.78

US3 – unskilled

0.67

NV1 – novice

0.89

NV2 – novice

0.88

NV3 – novice

0.91

EX1 – experienced

0.87

EX2 – experienced

0.97

EX3 – experienced

0.99

Computer program

1.00