From: Biomechanical testing of a unique built-in expandable anterior spinal internal fixation system
ROM (°) | Status | Expandable device (n = 6) | Z-plate (n = 6) | Kaneda (n = 6) | p |
---|---|---|---|---|---|
Flexion | Normal | 9.64 (0.81) | 9.98 (1.34) | 9.49 (0.94) | 0.717 |
Injury | 17.73 (3.67)a | 19.02 (3.12)a | 19.18 (3.21)a | 0.718 | |
Internal fixation | 5.72 (0.44)ab | 5.70 (0.50)ab | 5.63 (0.65)ab | 0.957 | |
Extension | Normal | 8.45 (0.52) | 8.45 (0.43) | 8.48 (0.44) | 0.994 |
Injury | 13.51 (2.21)a | 13.50 (2.53)a | 13.43 (2.25)a | 0.998 | |
Internal fixation | 5.25 (0.28)ab | 5.14 (0.31)ab | 5.12 (0.19)ab | 0.671 | |
Left bending | Normal | 8.71 (0.30) | 8.78 (0.27) | 8.55 (0.41) | 0.484 |
Injury | 18.07 (3.66)a | 18.23 (3.81)a | 18.87 (4.22)a | 0.932 | |
Internal fixation | 5.50 (0.21)ab | 5.37 (0.21)ab | 5.04 (0.13)*†ab | 0.002 | |
Right bending | Normal | 8.35 (0.37) | 8.47 (0.42) | 8.43 (0.39) | 0.869 |
Injury | 18.45 (3.72)a | 18.69 (3.80)a | 19.05 (2.54)a | 0.955 | |
Internal fixation | 5.01 (0.12)ab | 5.05 (0.14)ab | 5.04 (0.11)ab | 0.887 | |
Left axial rotation | Normal | 7.11 (0.40) | 6.97 (0.34) | 7.07 (0.28) | 0.769 |
Injury | 19.56 (2.98)a | 19.05 (3.22)a | 19.04 (3.26)a | 0.948 | |
Internal fixation | 4.53 (0.42)ab | 5.23 (0.30)*ab | 5.02 (0.15)*ab | 0.005 | |
Right axial rotation | Normal | 6.95 (0.35) | 6.91 (0.31) | 6.84 (0.44) | 0.883 |
Injury | 19.60 (3.18)a | 18.65 (3.41)a | 19.41 (3.35)a | 0.873 | |
Internal fixation | 4.49 (0.47)ab | 5.23 (0.20)*ab | 5.08 (0.17)*ab | 0.002 |