Skip to main content

Table 4 Fit and fill analysis between stem and endosteal canal

From: Design process of cementless femoral stem using a nonlinear three dimensional finite element analysis

Level

Fit (%)

Anteroposterior view

Mediolateral view

  

Femur (mm2)

Implant (mm2)

Fill (%)

Femur (mm2)

Implant (mm2)

Fill (%)

Proximal

42.89

873.42

843.29

96.55

1108.56

600.00

54.12

Middle

58.18

586.19

505.35

86.21

604.49

470.10

77.77

Distal

62.65

934.52

621.90

62.19

747.46

621.90

83.20

Total

53.68

-

-

83.10

-

-

71.70