Skip to main content

Table 2 Accuracy TSD - ADAPT vs. conventional technique - descriptive statistics

From: Improved lag screw positioning in the treatment of proximal femur fractures using a novel computer assisted surgery method: a cadaveric study

   Experience level  
Technique Descriptives Less experienced Experienced Overall
Conventional n 9 13 22
Median [mm] 4.40 4.40 4.40
Range [mm] 8.50 5.70 8.50
Interquartile range [mm] 3.60 4.10 3.70
ADAPT n 9 14 23
Median [mm] 5.00 4.60 5.00
Range [mm] 1.20 2.90 2.90
Interquartile range [mm] 0.85 1.35 1.20