Skip to main content

Table 1 Mean differences and 95% limits of agreement

From: New method for detection of complex 3D fracture motion - Verification of an optical motion analysis system for biomechanical studies

Force[N]

Δy [μm]

Δz[μm]

α [deg]

0

-5 [-45; 35]

-37 [-77; 3]

-0.05 [-0.11;0.01]

10

-7 [-48; 33]

-32 [-80; 16]

-0.04 [-0.13; 0.05]

20

-21 [-61; 19]

-27 [-83; 29]

-0.02 [-0.15; 0.10]

30

-37 [-77; 3]

-21 [-86; 44]

-0.01 [-0.17; 0.16]

40

-52 [-92; -12]

-15 [-89; 59]

0.01 [-0.20; 0.22]

50

-66 [-106; -25]

-9 [-93; 75]

0.03 [-0.23; 0.28]

60

-81 [-121; -41]

-2 [-97; 93]

0.04 [-0.26; 0.34]

70

-93 [-134; -53]

5 [-100; 111]

0.06 [-0.29; 0.42]

80

-106 [-146; -65]

13 [-105; 131]

0.09 [-0.32; 0.49]

90

-122 [-163; -82]

21 [-109; 150]

0.11 [-0.35; 0.57]

100

-138 [-179; -98]

27 [-112; 167]

0.13 [-0.38; 0.64]