Treatment of thoracolumbar burst fractures by short-segment pedicle screw fixation using a combination of two additional pedicle screws and vertebroplasty at the level of the fracture: a finite element analysis

Abstract

Background

Traditional one-above and one-below four-screw posterior short-segment instrumentation is used for unstable thoracolumbar burst fractures. However, this method has a high rate of implant failure and early loss of reduction. The purpose of this study was to use finite element (FE) analysis to determine the effect of treating thoracolumbar burst fractures by short-segment pedicle screw fixation using a combination of two additional pedicle screws and vertebroplasty at the level of the fracture.

Methods

An intact T11-L1 spine FE model was created from the computed tomography images of a male subject. Four fixation models with posterior fusion devices (pedicle screws, rods, cross-link) were established to simulate an unstable thoracolumbar fracture with different fusion surgeries: short-segment fixation with: 1) a link (S-L); 2) intermediate bilateral screws (S-I); 3) a link and calcium sulfate cement (S-L-C); 4) intermediate bilateral screws and calcium sulfate cement (S-I-C). Different loading conditions (flexion, extension, lateral bending, and axial rotation) were applied on the models and analyzed with a FE package. The range of motion (ROM), and the maximum value and distribution of the implant stress, and the stress in the facet joint, were compared between the intact and fixation models.

Results

The ROM in flexion, extension, axial rotation, and lateral bending was the smallest in the S-I-C model, followed by the S-I, S-L-C, and S-L models. Maximum von Mises stress values were larger under lateral bending and axial rotation loadings than under flexion and extension loading. High stress was concentrated at the crosslink and rod junctions. Maximal von Mises stress on the superior vertebral body for all loading conditions was larger than that on the inferior vertebral body. The maximal von Mises stress of the pedicle screws during all states of motion were 265.3 MPa in S-L fixation, 192.9 MPa in S-I fixation, 258.4 MPa in S-L-C fixation, and 162.3 MPa in S-I-C fixation.

Conclusions

Short-segment fixation with two intermediate pedicle screws together with calcium sulfate cement at the fractured vertebrae may provide a stiffer construct and less von Mises stress of the pedicle screws and rods as compared to other types of short-segment fixation.

Peer Review reports

Background

Burst fractures account for approximately 20% of thoracolumbar fractures, and occur due to an axial loading force that results in failure to support the anterior and middle column [1, 2]. Surgery is usually indicated when there is a severe deformity, and/or neurologic deficit. Whether anterior or posterior surgery is the most effective treatment for burst fractures remains controversial. Some authors advise anterior surgery to remove retropulsed fragments [3, 4], but posterior surgery is popular because it is an easier approach and allows clearance of the spinal canal by ligamentotaxis.

One-above and one-below posterior short-segment instrumentation with fusion has been widely used for unstable thoracolumbar burst fractures for the past 3 decades [5]. Pedicular instrumentation enables kyphotic correction, indirect reduction of canal encroachment, and early mobilization. However, this method has a high rate of implant failure, and early loss of reduction because of loss of anterior support [6]. Over the past decade, some studies have demonstrated that augmentation of the fractured vertebra with absorbable bone cement could enhance fracture union and prevent implant failure. Liao et al. [7] and Korovessis et al. [8] demonstrated that injectable calcium sulfate cement or injectable calcium phosphate cement used as a transpedicular grafting material in thoracolumbar fractures could obtain clinical and radiographic results equal to autogenous cancellous bone graft.

In recent years, biomechanical and clinical studies have suggested that two additional screws inside the fractured vertebra could improve stability and provide better kyphotic correction. In a biomechanical study, Norton et al. [9] showed that two additional screws in the fractured vertebra (a six-screw construct) of an unstable thoracolumbar burst fracture could increase the stiffness of the implant, and reduce stress on each pedicle screw, as compared to a four-screw construct. However, there have been no studies examining the effect of augmentation by a combination of screws and bone cement at the fractured vertebra, and comparing the effect of this fixation method with other types of posterior short-segment instrumentation for thoracolumbar burst fractures.

We hypothesize that posterior short-segment instrumentation with fractured vertebra augmentation by a combination of vertebroplasty and two additional screws can provide a stronger construct than other types of posterior short-segment instrumentation for thoracolumbar burst fractures. In the current study, we established a finite element (FE) model of thoracolumbar burst fractures, and four posterior short-segment fixation methods were tested. The purpose was to determine the most optimal methods for the treatment of thoracolumbar burst fractures.

Methods

Models

Four fixation models with posterior fusion devices (pedicle screws, rods, cross-link) were established to simulate an unstable thoracolumbar fracture with different fusion surgeries: short-segment fixation with: 1) a link (S-L); 2) intermediate bilateral screws (S-I); 3) a link and calcium sulfate cement (S-L-C); 4) intermediate bilateral screws and calcium sulfate cement (S-I-C). Radiographs representing these four types of fixation are shown in Fig. 1.

Fig. 1
figure 1

Four types of fixation. S-L: four screws with a link. S-I: six-screw construct. S-L-C: four screws with a link, and calcium sulfate bone cement vertebroplasty. S-I-C: six screws with calcium sulfate bone cement vertebroplasty

Establishment of FE model of an intact normal thoracolumbar spine

A FE model of an intact normal spine was developed based on 1-mm-interval cross-sectional computed tomography (CT) spine images of a 55-year-old male provided by the US National Library of Medicine (NLM; NIH, Bethesda, MD). The FE model of the thoracolumbar section from T10 to L2 was established as follows. 1) A series of CT scans were imported into Amira 4.1 software (Visage Imaging, Carlsbad, CA). The vertebral boundaries as regions of interest (ROI) were identified on multiple images to form a three-node triangular surface model. 2) The surface model was imported into SolidWorks (SOLIDWORKS Corporation, Boston, MA), to further reconstruct a three-dimensional solid model of the thoracolumbar spine. 3) The solid model of the thoracolumbar spine was imported into HyperWorks 10.0 (Altair Engineering, Inc., Troy, MI) to form an eight-node hexahedral FE model. Since CT scanning does not provide the soft tissue contours of the intervertebral discs, the geometric characteristics of intervertebral discs were created as previously described [10]. The volume ratio of the annulus fibrosus to the nucleus pulposus was set to 6:4, the thickness of the cortical bone was set to 1 mm, and the endplate was set to 0.5 mm [10]. 4) The final FE model was imported into Abaqus FE analysis software (Abaqus/CAE v.6.10; Simulia Corp., Providence, RI) for analysis (Fig. 2).

Fig. 2
figure 2

T11-L1 finite element model

Establishment of FE models of injured vertebra

In this study, the vertebral resection (corpectomy) method was used to simulate a vertebra fracture. Briefly, one-half of the sponge bone of the T12 vertebra was removed to weaken the vertebral strength; the structure of the posterior portion was reserved to establish a FE model of an unstable thoracolumbar fracture (Fig. 3).

Fig. 3
figure 3

T12 injured vertebrae finite element model

Establishment of solid internal fixator models

Pedicle screw system

The pedicle screw system in this study was modeled based on the SmartLoc Spinal Fixation System (A-SPINE Asia Co., Ltd.). Using physical measurements and 3D reverse engineering, a solid model was created using Solidworks. The length of the screw was 45 mm, the radius was 2.4 mm, and the cross-sectional area was 18.1 mm2. The length of the vertical connecting rod was 80 mm, the radius was 2 mm, and the cross-sectional area was 12.6 mm2. The horizontal connecting rod was 35 mm long, 7 mm wide, and 5 mm high. In order to shorten the FE analysis time, the screw thread part and the vertical connecting rod middle connection part of the structure were neglected. After the solid models of the implants were created, they were imported into HyperWorks to generate the FE meshes. The four-vertebral pedicle screw plus transverse connecting rod FE mesh model is shown in Fig. 4a, and the six-vertebral pedicle screw model is shown in Fig. 4b.

Fig. 4
figure 4

Finite element models of internal fixations. a: Four screws with a link. b: Six-screw construct

Four fixation models

The internal fixators were placed in the thoracolumbar models as described by Park et al. [11]. Four fixation models were used for comparison in this study: S-L, S-I, S-L-C, S-I-C. FE models of these four fixation types are shown in Fig. 5.

Fig. 5
figure 5

Four-screw fixation finite element model with T11-L1 spine segments. S-L, short-segment pedicle screw fixation with a link; S-I, short-segment pedicle screw fixation with two intermediate screws; S-C, short-segment pedicle screw fixation with cement vertebroplasty; S-I-C, short-segment pedicle screw fixation with two intermediate screws plus cement vertebroplasty

Material parameter settings

The material parameters (Young’s moduli and the Poisson’s ratios of cortical bone, cancellous bone, posterior bone element, endplate, and intervertebral disc) of the thoracolumbar models were set based on the work of Qiu et al. [12]. The models were assumed to be homogeneous, isotropic, and linearly elastic. Material parameters are summarized in Table 1.

Table 1 The material properties of the finite element model

Loading and boundary conditions, and friction coefficient settings

The loading and boundary conditions of the FE analysis were based on the work of Qiu et al. [13]. Flexion, extension, lateral bending, and axial rotation were applied to the rigid plane above the T11 vertebra, respectively, and the load was applied as a set in a progressive manner (2.5 Nm, 5.0 Nm, 7.5 Nm of bending moment), respectively, and all degrees of freedom on the bottom nodes of L1 were constrained. To simulate a real contact setting, the friction coefficient of the contact surface of the facet joint was set to 0.1. This study simulated bone fusion following surgery, therefore, the interfaces between the screw and vertebrae were set to be perfectly bonded.

Assessment indexes

By evaluating the range of motion (ROM) of T11-L1, the maximum von Mises stress on the pedicle screw root, and stress distribution of the pedicle screws and rods of the four fixation FE models under flexion, extension, left/right lateral bending, and left/right axial rotation, the biomechanical differences between the four fixation models could be compared.

Results

Validation of the intact T11-L1 FE model

After establishment of the normal T11-L1 FE model, moments of 7.5 Nm in six directions were obtained. The ROM of the T11-T12 segment were: flexion (3.0°), extension (3.1°), left bending (3.3°), right bending (3.8°), left rotation (2.2°), and right rotation (2.0°). The ROM of the T12-L1 segment were: flexion (3.3°), extension (3.7°), left bending (4.0°), right bending (3.6°), left rotation (1.3°), and right rotation (1.4°) (Table 2). The ROM results were similar to those reported by Panjabi et al. [14].

Table 2 Comparison between the present intact model and that of Panjabi et al. [12]

ROM of the four FE fixation models

The moments of 7.5 Nm in six directions were applied to the four fixation models. The ROM of the T11-L1 segment were: S-L group: flexion (1.03°), extension (1.05°), left bending (2.92°), right bending (2.80°), left rotation (2.60°), right rotation (2.60°); S-I group: flexion (0.95°), extension (0.98°), left bending (2.67°), right bending (2.54°), left rotation (2.32°), right rotation (2.32°); S-L-C group: flexion (0.98°), extension (1.00°), left bending (2.75°), right bending (2.61°), left rotation (2.46°), right rotation (2.46°); S-I-C group: flexion (0.87°), extension (0.92°), left bending (2.57°), right bending (2.51°), left rotation (2.23°), right rotation (2.23°) (Table 3).

Table 3 Range of motion of four fixations with different loading

Comparisons between the four fixation models and the intact model

The percentage of ROM of the four fixation models with respect to the intact model was compared in terms of the degree of ROM in each direction of movement. The percentage of ROM of the S-I-C group was lowest in all directions of motion (flexion: 13.2%; extension: 13.5%; left bending: 35.0%; right bending: 33.9%; left rotation: 64.6%; right rotation: 65.6%), indicating the stability was the best. The percentage ROM of the S-L group was the highest in all directions of motion (flexion: 15.6%; extension: 15.4%; left bending: 39.7%; right bending: 37.8%; left rotation: 75.3%; right rotation: 76.5%), indicating the worst stability. The S-I group ranked second, and the S-L-C group ranked third with respect to stability (Fig. 6).

Fig. 6
figure 6

Range of motion (ROM). Compared to the intact normal spine model, the four fixation models showed a decreased ROM in all directions. S-I-C fixation had the lowest ROM

Stress distribution of the pedicle instrumentation system

von Mises stress distribution analysis showed that high stresses were mainly concentrated on the screws during flexion and extension. High stress in lateral bending was mainly concentrated on the longitudinal connecting rod. High stress was also concentrated at the junction of the transverse connecting rod and the longitudinal connecting rod under a rotation load (Fig. 7).

Fig. 7
figure 7

Stress nephogram of the pedicle screws and rods with rotation stress

Maximum von Mises stress on the root of the pedicle screw

von Mises stresses of the pedicle screw neck of the S-L group were 30.9 MPa in flexion, 31.7 MPa in extension, 200.7 MPa in left lateral bending, 202.3 MPa in right lateral bending, 261.9 MPa in left rotation, and 265.3 MPa in right rotation. Stresses in the S-I group were 24.7 MPa in flexion, 25.7 MPa in extension, 141.8 MPa in left lateral bending, 144.3 MPa in right bending, 192.9 MPa in left rotation, and 191.2 MPa in right rotation. Stresses in the S-L-C group were 25.9 MPa in flexion, 26.1 MPa in extension, 156.5 MPa in left lateral bending, 258.4 MPa in right lateral bending, 200.4 MPa in left rotation, and 201.6 in right rotation. Stresses in the S-I-C group were 23.5 MPa in flexion, 24.6 MPa in extension, 128.6 MPa in left lateral bending, 130.2 MPa in right lateral bending, 162.3 MPa in left rotation, and 162.3 MPa in right rotation (Table 4).

Table 4 Maximum von Mises stress on the root of the pedicle screw

Discussion

Surgical management of a thoracolumbar burst fracture varies according to many factors. Fracture morphology, neurologic status, and surgeon preference all play major roles in deciding the surgical approach. An anterior approach to the fracture site can directly decompress the neural elements, and the reconstruction ca be performed by simultaneous iliac bone graft and plating. Hitchon et al. [15] and Sasso et al. [16] reported that the anterior approach was superior to the posterior approach in its ability to maintain the kyphotic correction at final follow-up, but both approaches achieved similar in clinical results. A major disadvantage of anterior surgery is potential donor site morbidity from harvesting iliac tri-cortical bone graft. In addition, the cost of the anterior approach is generally greater than that of the posterior approach. We usually use a posterior approach for the repair of thoracolumbar burst fractures, which is one of the reasons that led to the current study.

Traditional one-above and one-below short-segment posterior instrumentation has the potential for early implant failure and re-kyphosis due to lack of anterior support of the defect inside the injured vertebra. Several techniques, such as reinforcement with additional screws at the fractured level, or augmentation with any kind of bone cement to fill the defect of the fractured vertebra, have been proposed to improve the stability of the posterior instrumentation construct, and thus prevent implant failure and enhance fracture union [17,18,19,20]. However, no studies examined reinforcement of a fractured vertebra with a combination of bone cement with two additional fracture-level screws. Several biomechanical studies have suggested that reinforcement with fracture-level screws could improve the biomechanical stability of the construct [21,22,23]. Clinical studies have also suggested that reinforcement with additional screws at the fractured level can provide better kyphotic correction, more effectively restore fractured vertebra height, and allow earlier ambulation for patients with thoracolumbar burst fractures [24, 25]. On the contrary, authors have also claimed that a fractured vertebra augmented by absorbable bone cement followed by a posterior short-segment construct can provide satisfactory clinical results, with a low implant failure rate of 0% to 5% [19, 20]. Xu et al. [26] designed a FE model of a thoracolumbar burst fracture, and demonstrated that vertebroplasty inside the fractured vertebra can significantly reduce the stresses of the pedicle instrumentations and spine to prevent kyphotic correction loss and implant failure. In a clinical study, Liao et al. [27] showed that two screws placed inside the fractured vertebra with short-segment instrumentation was associated with shorter surgical time and less implant failure as compared to augmentation with injectable calcium sulphate/phosphate cement following posterior short-segment instrumentation.

In the current study, the four fixation models all demonstrated decreased ROM in all directions as compared with the intact spine model. ROM in flexion, extension, axial rotation, and lateral bending was the smallest in the S-I-C fixation model, followed by the S-I and S-L-C fixation models; the greatest ROM was seen in the S-L fixation model. The S-I-C and the S-I fixation model both contained six screws in the construct; the S-L-C and the S-L fixation only had four screws in the construct. Reinforcement with two additional screws at the fractured level can enhance biomechanical stability; and these results were similar to a previous FE study by Li et al. [28]. Two additional screws inside the fractured vertebra was also stronger than when bone cement was used inside the fractured level (S-I fixation model versus S-L-C fixation model). This result was similar to the clinical results demonstrated by Liao et al. [27].

The pedicle screw is considered to be the weakest point of the posterior fixation, and implant failure is usually caused by screw loosening or screw root breakage [29]. It has also been shown that the pedicle screws still continue to bear most of the load after fusion [30, 31]. To prevent early screw breakage, and to prolong the life of the screw, it is necessary to prevent the load on the screw from reaching the fatigue load. For this reason, we evaluated the maximum von Mises stress on the root of the pedicle screws in the four fixation models. A lower stress distribution around the root of the pedicle screw means the less chance of pedicle screw breakage. The maximum von Mises stress of the pedicle screws in the S-L group was higher than that in the other three fixation groups. In the S-I-C group, the maximum von Mises stress of the pedicle screw in all directions was the lowest. These results indicate that using two additional screws inside the fractured vertebra can reduce the stress on the pedicle screw root. Further augmentation with bone cement at the fractured level can further reduce the stress around the pedicle screw root, and further decrease the probability of screw failure. In addition, we found the pedicle screw root under lateral bending and axial rotation load sustained higher von Mises stresses, as finding similar to that of Xu et al. [32].

Injectable bone substitutes such as calcium sulfate, calcium phosphate cement, and hydroxyapatite cement are widely used for filling bone defects. These bone substitutes not only have osteoconductive ability to promote bone union, but also provide initial mechanical support. Evaniew et al. [33] used calcium sulfate/phosphate cement to manage patients with bone defects after curettage of primary tumor. Orsini et al. [34] demonstrated that calcium sulfate cement could promote new bone formation in a rabbit model of bone defects. Furthermore, Xu et al. [26] used a FE model to show that cement augmentation of a fractured vertebra could decrease the von Mises stress on the rods by 50% and on the screws by 40%. In the current study, however, the initial stability provided by bone cement inside the fractured vertebra did not achieve the stability provided by two additional screws. Nevertheless, an advantage of bone cements is that they can stimulate bone healing inside the vertebra, which screws cannot provide. Therefore, a combination of bone cement and two additional screws inside the fractured vertebra along with short-segment instrumentation may be an ideal surgical method for thoracolumbar burst fractures because provides greater initial stability, and also stimulates bone growth.

There are limitations of this study that should be considered. First, the FE spine models were reconstructed from data of a single patient, and thus is not representative of various ages or different sexes. Second, most cases of thoracolumbar burst fractures are associated with injury to the upper endplate and/or adjacent intervertebral disc. This was not addressed in the current models. Removal of the upper end-plate of the vertebrae in the FE model may better reflect the real injury pattern of a burst fracture. However, the main purpose of this study was to compare different fixation techniques, not different injuries. Third, because the current FE study only provided comparison data for the stability of different fixation methods, and the material representations of the biological structures were assumed to be linearly elastic, biomechanical studies of cadavers are necessary to validate the results for clinical practice. If the results are validated in cadaver studies, clinical cohort studies would be warranted.

Conclusion

Additional bilateral pedicle screws combined with cement augmentation vertebroplasty at the level of the fractured vertebra results in a stiffer construct and lower von Mises stress on the pedicle screws as compared with other types of posterior short-segment fixations. Four-screws plus a link resulted in the highest stress on the screws, suggesting this method would be associated with a higher incidence of implant failure and re-kyphosis. The S-I-C fixation was the strongest posterior short-segment fixation method for thoracolumbar burst fractures.

Abbreviations

FE:

Finite element

ROM:

Range of motion

Reference

  1. Benson DR, Burkus JK, Montesano PX, Sutherland TB, McLain RF. Unstable thoracolumbar and lumbar burst fractures treated with the AO fixateur interne. J Spinal Disord. 1992;5:335–43.

    Article  CAS  PubMed  Google Scholar 

  2. Crawford NR, Dickman CA. Construction of local vertebral coordinate systems using a digitizing probe. Tech Note Spine. 1997;22:559–63.

    Article  CAS  Google Scholar 

  3. Limb D, Shaw DL, Dickson RA. Neurological injury in thoracolumbar burst fractures. J Bone Joint Surg Br. 1995;77:774–7.

    CAS  PubMed  Google Scholar 

  4. Kaneda K, Taneichi H, Abumi K, Hashimoto T, Satoh S, Fujiya M. Anterior decompression and stabilization with the Kaneda device for thoracolumbar burst fractures associated with neurological deficits. J Bone Joint Surg Am. 1997;79:69–83.

    Article  CAS  PubMed  Google Scholar 

  5. Steffee AD, Biscup RS, Sitkowski DJ. Segmental spine plates with pedicle screw fixation. A new internal fixation device for disorders of the lumbar and thoracolumbar spine. Clin Orthop Relat Res. 1986;203:45–53.

    Google Scholar 

  6. Müller U, Berlemann U, Sledge J, Schwarzenbach O. Treatment of thoracolumbar burst fractures without neurologic deficit by indirect reduction and posterior instrumentation: bisegmental stabilization with monosegmental fusion. Eur Spine J. 1999;8:284–9.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Liao JC, Fan KF, Keorochana G, Chen WJ, Chen LH. Transpedicular grafting after short-segment pedicle instrumentation for thoracolumbar burst fracture: calcium sulfate cement versus autogenous iliac bone graft. Spine. 2010;35:1482–8.

    Article  PubMed  Google Scholar 

  8. Korovessis P, Hadjipavlou A, Repantis T. Minimal invasive short posterior instrumentation plus balloon kyphoplasty with calcium phosphate for burst and severe compression lumbar fractures. Spine. 2008;33:658–67.

    Article  PubMed  Google Scholar 

  9. Norton RP, Milne EL, Kaimrajh DN, Eismont FJ, Latta LL, Williams SK. Biomechanical analysis of four- versus six-screw constructs for short-segment pedicle screw and rod instrumentation of unstable thoracolumbar fractures. Spine J. 2014;14:1734–9. See comment in PubMed Commons below

    Article  PubMed  Google Scholar 

  10. Chosa E, Goto K, Totoribe K, Tajima N. Analysis of the effect of lumbar spine fusion on the superior adjacent intervertebral disk in the presence of disk degeneration, using the three-dimensional finite element method. J Spinal Disord Tech. 2004;17:134–9.

    Article  PubMed  Google Scholar 

  11. Park WM, Park YS, Kim K, Kim YH. Biomechanical comparison of instrumentation techniques in treatment of thoracolumbar burst fractures: a finite element analysis. J Orthop Sci. 2009;14:443–9.

    Article  PubMed  Google Scholar 

  12. Qiu TX, Tan KW, Lee VS, Teo EC. Investigation of thoracolumbar T12-L1 burst fracture mechanism using finite element method. Med Eng Phys. 2006;28:656–64.

    Article  PubMed  Google Scholar 

  13. Qiu TX, Teo EC, Zhang QH. Effect of bilateral facetectomy of thoracolumbar spine T11-L1 on spinal stability. Med Biol Eng Comput. 2006;44:363–70.

    Article  PubMed  Google Scholar 

  14. Panjabi MM, Oxland TR, Lin RM, McGowen TW. Thoracolumbar burst fracture. A biomechanical investigation of its multidirectional flexibility. Spine. 1994;19:578–85.

    Article  CAS  PubMed  Google Scholar 

  15. Hitchon PW, Torner J, Eichholz KM, Beeler SN. Comparison of anterolateral and posterior approaches in the management of thoracolumbar burst fractures. J Neurosurg Spine. 2006;5:117–25.

    Article  PubMed  Google Scholar 

  16. Sasso RC, Renkens K, Hanson D, Reilly T, McGuire RA Jr, Best NM. Unstable thoracolumbar burst fractures: anterior-only versus short-segment posterior fixation. J Spinal Disord Tech. 2006;19:242–8.

    Article  PubMed  Google Scholar 

  17. Guven O, Kocaoglu B, Bezer M, Aydin N, Nalbantoglu U. The use of screw at the fracture level in the treatment of thoracolumbar burst fractures. J Spinal Disord Tech. 2009;22:417–21.

    Article  PubMed  Google Scholar 

  18. Farrokhi MR, Razmkon A, Maghami Z, Nikoo Z. Inclusion of the fracture level in short segment fixation of thoracolumbar fractures. Eur Spine J. 2010;19:1651–6.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Marco RA, Meyer BC, Kushwaha VP. Thoracolumbar burst fractures treated with posterior decompression and pedicle screw instrumentation supplemented with balloon-assisted vertebroplasty and calcium phosphate reconstruction. Surgical technique. J Bone Joint Surg Am. 2010;92(Suppl 1 Pt 1):67–76.

    Article  PubMed  Google Scholar 

  20. Shen YX, Zhang P, Zhao JG, Xu W, Fan ZH, Lu ZF, et al. Pedicle screw instrumentation plus augmentation vertebroplasty using calcium sulfate for thoracolumbar burst fractures without neurologic deficits. Orthop Surg. 2011;3:1–6.

    Article  CAS  PubMed  Google Scholar 

  21. Anekstein Y, Brosh T, Mirovsky Y. Intermediate screws in short segment pedicular fixation for thoracic and lumbar fractures: a biomechanical study. J Spinal Disord Tech. 2007;20:72-77.

  22. Baaj AA, Reyes PM, Yaqoobi AS, Uribe JS, Vale FL, Theodore N, et al. Biomechanical advantage of the index-level pedicle screw in unstable thoracolumbar junction fractures. J Neurosurg Spine. 2011;14:192–7.

    Article  PubMed  Google Scholar 

  23. Wang H, Li C, Liu T, Zhao WD, Zhou Y. Biomechanical efficacy of monoaxial or polyaxial pedicle screw and additional screw insertion at the level of fracture, in lumbar burst fracture: an experimental study. Indian J Orthop. 2012;46:395–401.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Mahar A, Kim C, Wedemeyer M, Mitsunaga L, Odell T, Johnson B, et al. Short-segment fixation of lumbar burst fractures using pedicle fixation at the level of the fracture. Spine. 2007;32:1503–7.

    Article  PubMed  Google Scholar 

  25. Tian JW, Wang L, Xia T, Liu CY, Zhao QH, Dong SH. Posterior short-segmental fixation combined with intermediate screws vs conventional intersegmental fixation for monosegmental thoracolumbar fractures. Orthopedics. 2011;34:e389–96.

    Article  PubMed  Google Scholar 

  26. Xu G, Fu X, Du C, Ma J, Li Z, Ma X. Biomechanical effects of vertebroplasty on thoracolumbar burst fracture with transpedicular fixation: a finite element model analysis. Orthop Traumatol Surg Res. 2014;100:379–83.

    Article  CAS  PubMed  Google Scholar 

  27. Liao JC, Fan KF. Posterior short-segment fixation in thoracolumbar unstable burst fractures - Transpedicular grafting or six-screw construct? Clin Neurol Neurosurg. 2016;153:56–63.

    Article  PubMed  Google Scholar 

  28. Li C, Zhou Y, Wang H, Liu J, Xiang L. Treatment of unstable thoracolumbar fractures through short segment pedicle screw fixation techniques using pedicle fixation at the level of the fracture: a finite element analysis. PLoS One. 2014;9:e99156.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Moran JM, Berg WS, Berry JL, Geiger JM, Steffee AD. Transpedicular screw fixation. J Orthop Res. 1989;7:107–14.

    Article  CAS  PubMed  Google Scholar 

  30. Chen CS, Cheng CK, Liu CL. A biomechanical comparison of posterolateral fusion and posterior fusion in the lumbar spine. J Spinal Disord Tech. 2002;15:53–63.

    Article  PubMed  Google Scholar 

  31. Suda K, Ito M, Abumi K, Haba H, Taneichi H, Kaneda K. Radiological risk factors of pseudoarthrosis and/or instrument breakage after PLF with the pedicle screw system in isthmic spondylolisthesis. J Spinal Disord Tech. 2006;19:541–6.

    Article  PubMed  Google Scholar 

  32. Xu G, Fu X, Du C, Ma J, Li Z, Tian P, et al. Biomechanical comparison of mono-segment transpedicular fixation with short-segment fixation for treatment of thoracolumbar fractures: a finite element analysis. Proc Inst Mech Eng H. 2014;228:1005–13.

    Article  PubMed  Google Scholar 

  33. Evaniew N, Tan V, Parasu N, Jurriaans E, Finlay K, Deheshi B, et al. Use of a calcium sulfate-calcium phosphate synthetic bone graft composite in the surgical management of primary bone tumors. Orthopedics. 2013;36:e216–22.

    Article  PubMed  Google Scholar 

  34. Orsini G, Ricci J, Scarano A, Pecora G, Petrone G, Iezzi G, et al. Bone-defect healing with calcium-sulfate particles and cement: an experimental study in rabbit. J Biomed Mater Res B Appl Biomater. 2004;68:199–208. See comment in PubMed Commons below

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

This study was funded by the Chang Gung Memorial Hospital Research Fund. The grant number is CRRPG3E0151.

Availability of data and materials

The raw data belongs to our institution archive. All the necessary information is contained in the manuscript. We will be glad to share our data upon a reasonable request.

Author contributions

JCL provided the concept of the study and wrote the manuscript. WPC designed the study and took responsibility for revising the manuscript. HW designed the study and performed the anaalyses. All authors read and approved the final manuscript.

Competing interest

The authors declare that they have no competing interests.

Consent for publication

“Not applicable”

Ethics approval and consent to participate

This study was performed after obtaining ethics approval from Institute Review Board (IRB) of Chang Gung Memorial Hospital (IRB number: 105-6778C1). However, participant’s consent was not needed because this study did not involve any living person or animal body. Also, the license agreement to use images of Visible Human Data set for scientific research was granted by the Project officer at National Library of Medicine, U.S.A. on Nov. 17, 2015.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Weng-Pin Chen.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Liao, JC., Chen, WP. & Wang, H. Treatment of thoracolumbar burst fractures by short-segment pedicle screw fixation using a combination of two additional pedicle screws and vertebroplasty at the level of the fracture: a finite element analysis. BMC Musculoskelet Disord 18, 262 (2017). https://doi.org/10.1186/s12891-017-1623-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12891-017-1623-0

Keywords